Infant home respiratory monitoring using pulse oximetry

    Research output: Contribution to journalArticlepeer-review

    2 Downloads (Pure)

    Abstract

    Respiratory rate (RR) is a valuable early marker of illness in vulnerable infants, but current monitoring methods are unsuitable for sustained home use. We have demonstrated accurate measurement of RR from brief recordings of pulse oximeter plethysmogram (pleth) trace in full-term neonates in hospital. This study assessed the feasibility of this method in preterm infants during overnight recordings in the home. We collected simultaneous overnight SpO2, pleth and respiratory inductive plethysmography (RIP) on 24 preterm infants in the home. RR from pleth analysis was compared with RR from RIP bands; pleth quality was assessed by the presence of visible artefact. Median (range) RR from RIP and pleth were not significantly different at 42 (25-65) and 42 (25-64) breaths/min. Median (range) % of epochs rejected due to artefact was 20 (8-75) for pleth and 10 (3-53) for RIP. Our results suggest that home RR monitoring by pulse oximeter pleth signal is accurate and feasible.
    Original languageEnglish
    Pages (from-to)603-605
    JournalArchives of Disease in Childhood
    Volume103
    Early online date7 Jul 2016
    DOIs
    Publication statusPublished - 27 Sept 2018

    Bibliographical note

    Note: This work was supported by the National Institute for Health Research [grant number PB-PG-0610-22433].

    Keywords

    • Allied health professions and studies

    Fingerprint

    Dive into the research topics of 'Infant home respiratory monitoring using pulse oximetry'. Together they form a unique fingerprint.

    Cite this